• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Visual inspection of the returned product found that one of the batteries in the battery pack had leaked and crystalized inside the pack.Functional testing confirmed the device would not power on when connected to the battery pack returned with the device.After removing the faulty battery and replacing it with a new battery, the pulsavac powered on and functioned normally device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that the pulsa vac didn't work properly.The surgeon used another product.No, further information on what happened with the device/ lot number 64750040 / the incident occurred before the start of use.0-15 minute delay to the surgery to obtain alternate device.No harm / injury to the patient or operator.Investigation showed the battery pack had leaked and crystalized.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11732246
MDR Text Key247552337
Report Number0001526350-2021-00487
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375161
UDI-Public(01)00889024375161(17)230630(10)64750040
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot Number64750040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/27/2021
Initial Date FDA Received04/27/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-