• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. OSS AVL YOKE 12MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. OSS AVL YOKE 12MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Failure of Implant (1924); Pain (1994)
Event Date 08/14/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001825034-2021-01278, 0001825034-2021-01280, 0001825034-2021-01281, 0001825034-2021-01282, 0001825034-2021-01283, 0001825034-2021-01284.Medical product: oss avl tibial lock ring, item# 161073, lot# 466660.Oss avl mod tib plate oss tape, item# 161065, lot# 690080.Oss avl tib bearing 12mm, item# 161068, lot# 809890.Oss tib blk aug 10x63/67 univ, item# 150426, lot# 105590.Oss avl poly tib bushing set, item# 161071, lot# 490920.Oss poly femoral bushings, item# 150477, lot# 499870.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient was revised approximately five months¿ post-implantation due to locking nut fractured in two pieces and allowed piston to disengage from the tibial component.Attempts to obtain additional information have been made; however, no more information is available.
 
Manufacturer Narrative
Upon further investigation, it was determined that this item was reported in error.The locking ring fracture is what caused the yoke to disassociate from the tibial tray.Therefore, the locking ring is the only reportable item.The initial report was submitted in error and should be voided.
 
Event Description
Upon further investigation, it was determined that this item was reported in error.The locking ring fracture is what caused the yoke to disassociate from the tibial tray.Therefore, the locking ring is the only reportable item.The initial report was submitted in error and should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSS AVL YOKE 12MM
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11739289
MDR Text Key247765525
Report Number0001825034-2021-01279
Device Sequence Number1
Product Code KRO
UDI-Device Identifier00880304456143
UDI-Public(01)00880304456143(17)281101(10)382130
Combination Product (y/n)N
PMA/PMN Number
K051570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,health p
Type of Report Initial,Followup
Report Date 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number161075
Device Lot Number382130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2021
Initial Date FDA Received04/28/2021
Supplement Dates Manufacturer Received05/19/2021
Supplement Dates FDA Received05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10
Patient Outcome(s) Hospitalization; Required Intervention;
-
-