The electronic log file was available for investigation.The case in question could be reconstructed by means of the information stored therein and no indications for a device malfunction were found: the case in question was started in man/spont at 9:59am and continued in volume control.The last system test was successfully performed at 1:46am at the same day.The ventilation was subsequently inconspicuous and stable, the measured peep corresponded to the user setting, thus no peep/cpap high alarms were generated.It could be understood that there were repeated changes of therapy (between man/spont and volume control), most likely for therapeutic reasons.There were also repeated changes between low and high peep levels, probably for lung recruitment.Beginning at approximately 3:00pm, there were repeated significant increases in peak and plateau pressures (from approximately 20 to 40 hpa) while peep remained constant (set at 5 hpa at this time).The vte showed a reduction of about 50% during this phase.According to specification, the device repeatedly alarmed pressure high and volume not attained due to the set pressure limit.At 3:46pm, the user switched to standby.At this time, as further reported, the breathing system, tubing, and filter were changed while the patient was ventilated with an oxylog device.After successfully performing a system test, the therapy was continued from 4:16pm with the perseus and ventilation was subsequently stable and unremarkable again without further problems until the device was switched to standby at 5:21pm.Finally, the reported symptom could be partially reproduced based on the log analysis.Immediately prior to the change of the breathing system, there was an increase in peak and plateau pressures with peep according to the setting but reduced expiratory tidal volumes.This indicates a blocked filter or increased resistance in the expiratory path.There were no indications for a technical device malfunction found and the measured peep pressure corresponded to the set value at all times.Obviously, however, there was a significant delay in pressure relief on the patient side, resulting in an increase in pressure.The integrated pressure-, volume- and patient gas monitoring ensures that deviations from set or expected parameters are obvious and that alarms are given according to the alarm limits adjusted by the user.The device alarmed as specified.
|