MAUDE Adverse Event Report: SYNTHES GMBH T-PAL TRIAL SPACER 10MM X 28MM 7MM HEIGHT; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH T-PAL TRIAL SPACER 10MM X 28MM 7MM HEIGHT; TEMPLATE Back to Search Results
Catalog Number 03.812.307
Device Problem Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.The investigation could not be completed, no conclusion could be drawn at the time of filing this report.The complaint condition that the proximal ball of the trial implant is broken could be confirmed according to the received pictures.The applicator outer shaft is not visible on the received pictures.Therefore, the complaint condition of this instrument will be rated as n/a.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part: 03.812.307, lot: 8814473, manufacturing site: (b)(4).Release to warehouse date: apr.04, 2014.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent for a surgery.During the surgery, the surgeon used the trial with the applicator.While trying to remove the trial implant from the disk space the trial detached from the applicator.The surgery was completed successfully with 5 minutes delay.There were no patient consequences are reported.This complaint involves two (2) devices.This report is for (1) t-pal trial spacer 10mm x 28mm 7mm height.This report is 1 of 3 (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PAL TRIAL SPACER 10MM X 28MM 7MM HEIGHT
Type of Device
TEMPLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key11748830
MDR Text Key249748007
Report Number8030965-2021-03443
Device Sequence Number1
Product Code HWT
UDI-Device Identifier07611819418974
UDI-Public07611819418974
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.812.307
Device Lot Number8814473
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/07/2021
Initial Date FDA Received04/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-