Model Number SIGTRS45AMT
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2021
Event Type  malfunction  
Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the device was found to be partially fired and a damage to the cutting edge of the knife blade was observed.It was reported that after firing, the reinforcement material did not release from the device as expected.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during partial right pulmonary resection, when reload stapled the pulmonary parenchyma, the suture on the tip of the reload could not be cut, so the reinforcement sheet became unable to be removed from the pulmonary parenchyma.The surgeon cut the reinforcement sheet with scissors.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key11757616
MDR Text Key248441244
Report Number1219930-2021-01717
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884521543645
UDI-Public10884521543645
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIGTRS45AMT
Device Catalogue NumberSIGTRS45AMT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/13/2021
Initial Date FDA Received04/30/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-