• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that during a routine inspection, the light source was not switching on.No case was involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Factors unrelated to the manufacturing and design of the device which could have contributed to the reported event, include a defective power supply/ballast or other related electrical components.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future occurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE 500XL XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11758660
MDR Text Key250413648
Report Number1643264-2021-01641
Device Sequence Number1
Product Code FFS
UDI-Device Identifier03596010643070
UDI-Public03596010643070
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200568
Device Catalogue Number72200568
Device Lot NumberBBR3560
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/12/2021
Initial Date FDA Received05/01/2021
Supplement Dates Manufacturer Received06/02/2021
Supplement Dates FDA Received06/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-