In troubleshooting the issue with olympus technical support via the phone, the reporter stated the device would be sent in for evaluation/repair.The subject device has been received and is currently in the evaluation process.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
This supplemental report is being submitted to provide the results of the device evaluation and legal manufacturer¿s investigation.The device was returned to an olympus service center for evaluation and the reported issue was not confirmed.However, the evaluation found that the connection had heavy dust/debris blocking the scope socket pins which caused a b30 error.In addition, the device had a non-olympus lamp with over five hundred (500) hours of use and was out of specification.The legal manufacturer performed an investigation.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The root cause of the reported issue could not be conclusively specified.Regarding the issue of the clv-190 light source being shut down, as the issue could not be reproduced by olympus, it was likely that there was a factor such as the capacity of the outlet at the facility was smaller than the total electric capacity of the connected equipment.It was likely that error b30 occurred because dust and dust blocked the pins of the scope socket.The ifu (instruction for use) describes the following: 3.8 connection to the ac mains power supply be sure to connect the power plug securely to prevent unintentional disconnection during use.The equipment will not function.Do not extend a single wall mains outlet into multiple outlets for simultaneous.Malfunction of the equipment may result.Confirm that the hospital-grade wall mains outlet to which the light source is connected has adequate electrical capacity that is larger than the total power consumption of all connected equipment.If the capacity is insufficient, a fire can result, or the circuit breaker may trip, and it may turn off not only the light source but also all other equipment connected to the same power circuit.
|