Model Number 15
Device Problem Failure to Analyze Signal (1539)
Patient Problem Respiratory Failure (2484)
Event Date 04/05/2021
Event Type  malfunction  
Manufacturer Narrative
Physio-control contacted the customer to request additional information on the patient.The customer provided physio-control with all the available patient information.Patient fields in which information is not provided were intentionally left blank.Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the therapy connector of the device and then observed proper device operation through functional and performance testing.The device was returned to the customer for use.
 
Event Description
The customer contacted physio-control to report that their device paddles lead ecg displayed severe artifact.In this state, the device may not have the ability to function in aed mode, and defibrillation therapy may be delayed or prevented from being delivered to the patient if needed.The customer did advise that ecg was obtained through the patient ecg cable.The patient was unconscious and not breathing, resuscitation was done at the scene, but no return of spontaneous breathing or heartbeat.The patient did not survive, however the customer a health care provider confirmed that the device use did not contribute to the patient outcome.The patient did not require defibrillation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key11764163
MDR Text Key248668911
Report Number0003015876-2021-00979
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873884331
UDI-Public00883873884331
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-001373
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/15/2021
Initial Date FDA Received05/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age69 YR
Patient Weight45
-
-