• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2021
Event Type  malfunction  
Manufacturer Narrative
During troubleshoot via telephone, the clinical engineer technician reported that the filters were dirty and was causing the unit to overheat.The technician was able to correct the issue by cleaning the filters.And are no longer experiencing an e102 error.The investigation is ongoing; therefore, the root cause of the reported malfunction cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
During a procedure, the evis exera iii xenon light source was overheating and getting an e102 lamp outage failure error message and the light source kept going into spare lamp mode.There was no delay and the procedure was completed using the same device.There was no patient injury or harm was reported.
 
Manufacturer Narrative
The legal manufacturer performed the device history records for this device and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The investigation was completed by the legal manufacturer and determined that there is no manufacturing, material or processing related cause for this failure mode.However, since the reported e102 error did not occur after the filter and dust inside the device were cleaned, the legal manufacturer determined that the e102 error occurred due to dust accumulated inside the subject device.This increased the temperature inside the device, so the safety unit was activated.Then, e102 was displayed without turning on the main lamp, and the lamp was switched to the emergency lamp.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: the legal manufacturer will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11766740
MDR Text Key275259548
Report Number8010047-2021-05771
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/09/2021
Initial Date FDA Received05/03/2021
Supplement Dates Manufacturer Received05/31/2021
Supplement Dates FDA Received06/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-