It was reported an issue with dafilon suture.The client reports that in the process of suturing child with trauma, the suture was disconnected from the needle.Doctors immediately exchanged for new sutures and took off and wore sterile gloves again.Delayed processing time, child crying irritable.No further information has been received.
|
Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical's warehouse.We have not received any samples.Without closed and/or defective samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done, and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyze it.Please note that when no samples are received our analysis is very limited.We regret any inconvenience this issue may have caused and thank you for your collaboration.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|