DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3013-55 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse confirmed the issue.The fse replaced the motor controller pcb.The fse completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp is working in satisfactory condition.
|
|
Event Description
|
It was reported that during a routine check the cs100 intra-aortic balloon pump (iabp) showed electrical test failure code # 50.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
In addition to the repair information provided in medwatch report # 2249723-2021-00926, the iabp was returned to the customer and cleared for clinical use.Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period may-2019 through apr-2021 was reviewed.Run rule was triggered for the month of apr 2021.The trigger is addressed under capa request which is now closed.
|
|
Search Alerts/Recalls
|
|
|