• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 10CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 10CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66000041
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2021
Event Type  malfunction  
Event Description
It was reported that during set up or inspection the opsite flexifix 10cmx10m ctn 1 was creased.The procedure was completed using the same device.No case involved.
 
Manufacturer Narrative
H3, h6: the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review found a small number of similar instances of the reported event.There is nothing to indicate that this is outside of acceptable rates of occurrence.The device was not used for treatment and was returned for analysis.The roll was creased and was coming away from the carrier paper at one side, thus establishing a relationship between the event and the device.It has not been possible to identify a definitive root cause.Probable root cause is damaged caused during the packing of the roll within the carton during manufacture.Quality checks are in place to prevent damaged goods leaving the production site and so this may be as a result of a missed quality check.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIFIX 10CMX10M CTN 1
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11773627
MDR Text Key249014829
Report Number8043484-2021-01004
Device Sequence Number1
Product Code KGX
UDI-Device Identifier05000223411473
UDI-Public5000223411473
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2023
Device Catalogue Number66000041
Device Lot Number2037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Initial Date Manufacturer Received 04/12/2021
Initial Date FDA Received05/04/2021
Supplement Dates Manufacturer Received10/13/2021
Supplement Dates FDA Received10/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-