• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2021
Event Type  malfunction  
Manufacturer Narrative
Per data log analysis, there was no indication of an issue in the log.The user facility sent a video demonstrating the sound coming from the fraction of inspired oxygen (fio2) stepper motor.After review of the video, the sound did not seem unusual.
 
Event Description
It was reported that prior to use of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) was making an unusual sound and the stepper motor was thought to be malfunctioning.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint is not verifiable.Multiple diligence attempts for part return were unsuccessful so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
The reported complaint was not confirmed.During laboratory analysis, the product surveillance technician (pst) performed verification testing without any observation of the electronic patient gas system (epgs) making any unusual sounds.Both stepper motors sounded normal.The epgs met specifications.The service repair technician (srt) did not duplicate the reported complaint.The epgs was powered on and the srt waited 15 minutes for it to warmup.The air and oxygen (o2) regulator were set to 50 pounds per square inch (psi) and the o2 analyzer calibrated successfully.There was no observation of the epgs making an unusual sound.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11780275
MDR Text Key252014074
Report Number1828100-2021-00164
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/12/2021
Initial Date FDA Received05/05/2021
Supplement Dates Manufacturer Received05/14/2021
07/27/2021
Supplement Dates FDA Received06/08/2021
07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-