H10: manufacturing review: a complaint history review was performed.This is the first complaint reported to date for this product and lot, therefore a device history record review is not required.Investigation summary: the sample was not returned for evaluation, an electronic video was provided for review.The investigation is confirmed for the reported stretched and material protrusion issues as ballooning was noted upon infusion on the blue extension leg of the catheter just distal to the hub in the provided video.However, the investigation is inconclusive for the reported failure to aspirate as the exact circumstances at the time of the reported event are unknown.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 03/2021), g3, h6(method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|