• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 EQUISTREAM SPLIT TIP; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 EQUISTREAM SPLIT TIP; DIALYSIS CATHETER Back to Search Results
Catalog Number 6903190
Device Problems Stretched (1601); Suction Problem (2170); Deformation Due to Compressive Stress (2889); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 03/2021).
 
Event Description
It was reported that during treatment the catheter of the device allegedly kinked.It was further reported that the device allegedly failed to aspirate.There was no reported patient injury.
 
Event Description
It was reported that during treatment the catheter of the device allegedly had bulging on the extension leg.It was further reported that the device allegedly failed to aspirate.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported to date for this product and lot, therefore a device history record review is not required.Investigation summary: the sample was not returned for evaluation, an electronic video was provided for review.The investigation is confirmed for the reported stretched and material protrusion issues as ballooning was noted upon infusion on the blue extension leg of the catheter just distal to the hub in the provided video.However, the investigation is inconclusive for the reported failure to aspirate as the exact circumstances at the time of the reported event are unknown.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 03/2021), g3, h6(method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUISTREAM SPLIT TIP
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11782442
MDR Text Key249994288
Report Number3006260740-2021-01681
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
PMA/PMN Number
K090101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6903190
Device Lot NumberREDR3726
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/10/2021
Initial Date FDA Received05/06/2021
Supplement Dates Manufacturer Received08/22/2021
Supplement Dates FDA Received08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age32 YR
Patient Weight65
-
-