• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC. COOK FEMORAL A LINE; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC. COOK FEMORAL A LINE; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 04/26/2021
Event Type  malfunction  
Event Description
Femoral aline noted to be leaking profusely after pt.Supine.Dressing taken down and blood noted to be spurting from a spot in the aline catheter.Aline removed and catheter and tubing piece changed.Pt was very sick and a 3rd pressor was being initiated.We had to use nibp for pressor titration which was suboptimal.A-line tubing taken to risk.Unsure if line was punctured when being placed by md.Patient was being prone and supine.If line was punctured with suture while placing it may not have been noticed initially but with turning, wonder if it pulled out a little and then leak was seen.This was a femoral line and it was very difficult to place.They were unable to place a radial, so femoral done with much difficulty.Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK FEMORAL A LINE
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC.
bloomington IN 47402
MDR Report Key11783054
MDR Text Key249499757
Report NumberMW5101194
Device Sequence Number1
Product Code DQO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Type of Report Initial
Report Date 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/05/2021
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age63 YR
-
-