Qn #(b)(4).The customer reported the tubing burst during use.The customer returned one nebulizer kit which includes a jet, jar, cap, tubing, and mask.The returned sample was visually examined with and without magnification.Visual examination of the returned tubing revealed the tubing had burst.No other defects or anomalies were observed.A dimensional inspection test was performed on the wall thickness of the tubing, and no dimensional issues were found.The affected part of the tubing was cut and a new connector was installed on the tubing in order to perform functional testing.No issues were encountered and no leaks were found.The tubing is designed to allow oxygen flow even when there is an obstruction in the tubing.Based on the visual exam, the reported complaint was confirmed.Although the complaint was confirmed, a root cause for the issue could not be identified.There was no evidence to suggest a manufacturing related issue.
|