MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Model Number 0620040610
Device Problems Decrease in Pressure (1490); Appropriate Term/Code Not Available (3191)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that there was loss of insufflation during procedure.
 
Event Description
It was reported that there was loss of insufflation during procedure.
 
Manufacturer Narrative
Alleged failure: i have an account struggling with some pneumosure¿s.They¿re not able to sustain their pressure in a laparoscopic case.They are in high flow, they had their set pressure at 15.I had them increase up to 25mmhg and are still struggling.They¿ll fall to 7, over pressure and vary widely in a short period of time.They¿re using bottle gas, the gas tank icon is green.They are using 2x 5mm ports.They have now used multiple insufflators in this same case with the same issues.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.There is no probable root cause in relation to the complaint unit as there were no errors during the functional testing of the unit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11787548
MDR Text Key251918680
Report Number0002936485-2021-00246
Device Sequence Number1
Product Code HIF
UDI-Device Identifier07613327063295
UDI-Public07613327063295
Combination Product (y/n)N
PMA/PMN Number
K063367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0620040610
Device Catalogue Number0620040610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2021
Initial Date Manufacturer Received 04/07/2021
Initial Date FDA Received05/06/2021
Supplement Dates Manufacturer Received04/07/2021
Supplement Dates FDA Received08/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-