• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955558
Device Problems Use of Incorrect Control/Treatment Settings (1126); Improper or Incorrect Procedure or Method (2017)
Patient Problem Hypovolemia (2243)
Event Date 04/13/2021
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that at an unknown time during continuous renal replacement therapy using a prismax machine, the nurse mistakenly entered the pre blood pump flow rate into patient fluid removal and removed 1200 ml of fluid from the patient.The patient became hypovolemic and required a bolus of plasmalyte 500mls and 750mls albumin to compensate for the excessive fluid removal.The patient did not suffer any other adverse event.No additional information is available.
 
Manufacturer Narrative
Additional information: corrected information added to h6- device code.Code a2303 was added to capture user error.H10: a device history review revealed no issues that could have caused or contributed to the reported issue.The device was not returned; therefore, a device analysis could not be completed, however, based on the reported event description, the reported condition was verified.The cause was design related.This issue is being further investigated.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11790823
MDR Text Key249479515
Report Number1416980-2021-02730
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number955558
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/14/2021
Initial Date FDA Received05/07/2021
Supplement Dates Manufacturer Received06/24/2021
Supplement Dates FDA Received06/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
-
-