BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 2C8750
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a clearlink system y-type blood/solution set was cut in half.The was identified when the package was opened.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
D4: the lot # is dr20k14039, previously submitted as asku.D4: the udi # is (b)(4), previously submitted as ni.H10: the device was received for evaluation.A visual inspection was performed using the naked eye which observed a transversal cut in the tubing.The reported condition was verified.The cause of the condition was caused by the packaging machine during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11793803
MDR Text Key251285775
Report Number1416980-2021-02756
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)Y
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8750
Device Lot NumberDR20K14039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Initial Date Manufacturer Received 04/14/2021
Initial Date FDA Received05/07/2021
Supplement Dates Manufacturer Received06/16/2021
Supplement Dates FDA Received06/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-