MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II CR LOCK FEM IMPLANT IMPACTOR; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II CR LOCK FEM IMPLANT IMPACTOR; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74011711
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/15/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka surgery, the jrny ii cr lock fem implant impactor would not unlock the femoral component by hand while implanting the femoral component.The procedure was completed, with a delay of fewer than 30 min, using the same device.Patient was not harmed.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.Therefore, a product analysis could not be performed at this time and the reported event could not be confirmed.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR LOCK FEM IMPLANT IMPACTOR
Type of Device
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11794866
MDR Text Key249666605
Report Number1020279-2021-03887
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556422892
UDI-Public00885556422892
Combination Product (y/n)N
PMA/PMN Number
K121443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74011711
Device Catalogue Number74011711
Device Lot Number19CBG0105
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/15/2021
Initial Date FDA Received05/07/2021
Supplement Dates Manufacturer Received06/15/2021
Supplement Dates FDA Received06/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-