• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AEROGEN LIMITED AEROGEN SOLO CONTINUOUS NEBULIZATION TUBE SET; NEBULIZER (DIRECT PATIENT INTERFACE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AEROGEN LIMITED AEROGEN SOLO CONTINUOUS NEBULIZATION TUBE SET; NEBULIZER (DIRECT PATIENT INTERFACE) Back to Search Results
Model Number 90-066
Device Problems Product Quality Problem (1506); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Insufficient Information (4580)
Event Date 04/13/2021
Event Type  malfunction  
Event Description
Sbar: aerogen continuous inhalation medication delivery situation.Since (b)(6) of 2020 we have been placed on allocation for aerogen syringes due to covid-19.We have also had problems with the proprietary tubing from aerogen, as it is sometimes too long causing it to rest on the mesh and impeded nebulization.Background: (b)(6) of 2017, we were informed that aerogen solo luer connector was no longer going to be produced after (b)(6) 2017.We were given two options: we could go to a syringe pump system or a cme pump system.We decided on the cme method because we could continue to use the iv bags with the addition of the fused luer tubing.Additionally, the aerogen syringe is not sterile and cannot go under the pharmacy hood for compounding medication.(b)(6) 2019, we started using the cme pump.(b)(6) 2020, we were notified that cme pumps were recalled, forcing us to go to the syringe pump delivery system.(b)(6) 2020, we went back to using the single luer and the sigma spectrum pump until we could develop education and obtain syringe pumps.On (b)(6) 2021 we rolled out the medfusion syringe pump.That day respiratory care leads were made aware of the allocation that tri-anim had placed us on.That allocation was for 15 syringes a month.This was based on our other facility's usage since we had not started with the syringe pump yet.(b)(6) 2021, we reached out to get allocation increased to no avail.(b)(6) 2021, allotment was increased to 200 syringes a month.Assessment: current state we may or may not have enough syringe allotment for our hospitals' needs.If we look at our usage over the last six months, we would have averaged 268 syringes per month.Aerogen is the only device on the market that allows for medications to be aerosolized throughout the lung while not adding any additional flow to the patient's vent circuit.Because of their monopoly on this market; we are tied to using their proprietary supplies for delivery.Recommendation: we have developed a way utilizing non-proprietary iv tubing and a spike for delivering continuous nebulization using the aerogen nebulizers.We would not have to use their syringe or feedline with luer.Which would alleviate the continued issues we have with allocation as well as disposables failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEROGEN SOLO CONTINUOUS NEBULIZATION TUBE SET
Type of Device
NEBULIZER (DIRECT PATIENT INTERFACE)
Manufacturer (Section D)
AEROGEN LIMITED
2500 bee cave road
bldg 1, suite 300
austin TX 78746
MDR Report Key11796681
MDR Text Key249704575
Report Number11796681
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number90-066
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/19/2021
Event Location Hospital
Date Report to Manufacturer05/10/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-