H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection was conducted and confirmed the tech inserted the wrong capture stick into the lag screw driver and now it¿s stuck on the lag/comp screw kit 90/85 and won¿t come out.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|