• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71173377
Device Problems Material Twisted/Bent (2981); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/16/2021
Event Type  malfunction  
Event Description
It was reported that, during inspection or set up for a tka surgery, six (6) rdce frcps w/ pts brd were sticking and bent at the tips.The procedure was successfully completed without delay using a smith and nephew back up device.No other complications were reported.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RDCE FRCPS W/ PTS BRD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11803393
MDR Text Key249984702
Report Number1020279-2021-03964
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010519535
UDI-Public03596010519535
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71173377
Device Catalogue Number71173377
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/16/2021
Initial Date FDA Received05/10/2021
Supplement Dates Manufacturer Received06/03/2021
Supplement Dates FDA Received06/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-