DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Complete Blockage (1094); Gas Output Problem (1266); Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2021
Event Type  malfunction  
Manufacturer Narrative
The log file was analyzed whereby it was determined that the device passed the automatic power-on self-test (post) in the morning of the date of event without deviations.The respective surgical procedure ran stable and unremarkable for approx.30 minutes until the self-monitoring function detected a communication problem between the cpu board that controls the gas dosage and the user interface.In a situation where ventilator and gas mixer enter a fail state simultaneously the supervisor software routine is designed to switch-off both subsystems.The device went into the so-called "safety mode" and alerted the user to this condition by means of a corresponding alarm.If such deviation occurs during use the user has to set up an adequate oxygen and a-gas flow manually to perform manual ventilation with the in-built breathing bag; monitoring functionalities of the device are not affected by this kind of error condition.The procedure how to establish the emergency gas supply is laid down in the ifu.Dräger knows a certain number of similar events - none of these events could be traced back to a clear root cause.The fact that the identical type of board is used in the workstation twice and does not exhibit malfunctions in the second application periphery makes a general design issue rather unlikely.A reasonable explanation would be electrostatic discharge of the user during interaction with the device or any other kind of electromagnetic disturbance that exceeds the immunity barriers of the device.The primus was developed in compliance to the requirements of iec 60601-1-2.The number of similar cases is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that the device failed during use accompanied by a gas mixer failure alarm.No patient consequences have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key11804472
MDR Text Key252435357
Report Number9611500-2021-00202
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/29/2021
Initial Date FDA Received05/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-