SMITH & NEPHEW, INC. GII OVAL PATELLA 5 PT. SZ GDE; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Model Number 71440842 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, two (2) gii oval patella 5 pt.Sz gde were broken.It is unknown when occurred and how was the procedure finished.There was no surgical delay.Patient was not harmed.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
H6:the associated device, intended to use in treatment, was returned for evaluation.A visual inspection of the returned gii oval patella 5 pt.Sz gde confirms it is broken on one end.The broken piece was returned with the device.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|