|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/14/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that a ventilator experienced anomalies in numerical values.The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.The international field service engineer (fse) reported that a ventilator experienced abnormal average machine pressure values.The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.There was no delay to therapy reported.The device was evaluated by the fse who confirmed and duplicated the reported issue.The fse replaced the data acquisition board.The unit was functionally tested and successfully passed all testing.The unit was returned to service.No other anomaly was reported.
|
|
Manufacturer Narrative
|
G4:(b)(6)2021 g5:510k: k102985 b4:24jun2021 the visual inspection of this customer returned data acquisition (da) printed circuit board assembly (pcba) revealed no anomalies.A failure investigation (fi) technician installed the data acquisition (da) printed circuit board assembly (pcba) into a fi ventilator to duplicate the reported issue of abnormal proximal pressure values.Fi technician identified that the abnormal proximal pressure values could not be verified or reproduced.No fault was found.
|
|
Search Alerts/Recalls
|
|
|