THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 102956
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the centrimag motor had a bad set screw.The motor was sent in for repair.No additional information was provided.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of a damaged motor set screw was not confirmed as the centrimag motor (serial number: (b)(6)) was not returned for analysis and no photos were submitted for review.Information regarding the centrimag motor¿s return status was unable to be provided.The root cause of the reported event could not be conclusively determined through this analysis.The centrimag motor will be re-evaluated in the event that it is returned for analysis.The device history records were reviewed and the records revealed that the centrimag motor, serial number (b)(6), was manufactured in accordance with manufacturing and quality assurance specifications.The centrimag motor was shipped to the customer on 20jun2013.The 2nd generation centrimag system operating manual section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.Centrimag motor instructions for use instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key11821569
MDR Text Key251242758
Report Number3003306248-2021-01120
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140078
UDI-Public07640135140078
Combination Product (y/n)N
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number102956
Device Catalogue Number102956
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/22/2021
Initial Date FDA Received05/13/2021
Supplement Dates Manufacturer Received07/07/2021
Supplement Dates FDA Received07/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-