ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VICTORY XL DR; IMPLANTABLE PACEMAKER PULSE GENERATOR
|
Back to Search Results |
|
Model Number 5816 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/15/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during initial implant of a pacemaker, the physician was unable to insert a lead into the header of the device.The physician exchanged the pacemaker and the new device was successfully implanted.The patient was stable throughout.
|
|
Manufacturer Narrative
|
The reported field event of failure to connect leads to header was not confirmed in the laboratory.The device was tested on the bench and using automated testing equipment, and no anomalies were found.
|
|