• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL MEDFUSION; SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL MEDFUSION; SYRINGE Back to Search Results
Model Number 4000
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation completed on a smiths medical syringe infusion pumps|medfusion 4000 pumps complain of flash memory post error code was revealed in the event hisotry log as well as during testing.This was isolated to the main pc board.The cause was believed to be crc failure found in the program memory during power up self testing.Action was taken to replace the main pcb board.Device then passed all funcitonal testing.
 
Event Description
Information received a smiths medical syringe infusion pumps|medfusion 4000 pumps had system failure and memory post.No patient adverse events reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDFUSION
Type of Device
SYRINGE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11828914
MDR Text Key250798508
Report Number3012307300-2021-04371
Device Sequence Number1
Product Code MRZ
UDI-Device Identifier10610586043567
UDI-Public10610586043567
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4000
Device Catalogue Number4000-0105-51
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/16/2021
Initial Date FDA Received05/14/2021
Supplement Dates Manufacturer Received06/14/2023
Supplement Dates FDA Received06/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-