• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL MEDFUSION; SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL MEDFUSION; SYRINGE Back to Search Results
Model Number 4000
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation completed on a smiths medical syringe infusion pumps|medfusion 4000 pumps complain of base hardwood failure was revealed in the event hisotry log and confimred during powering the device up.The cause is multiple conponants of interconnect board and battery.Action was taken to replace the interconnect board and battery.
 
Event Description
Information received a smiths medical syringe infusion pumps|medfusion 4000 pumps alarmed base hardware failure.No patient adverse events reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDFUSION
Type of Device
SYRINGE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11829090
MDR Text Key250798002
Report Number3012307300-2021-04373
Device Sequence Number1
Product Code MRZ
UDI-Device Identifier10610586043567
UDI-Public10610586043567
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number4000
Device Catalogue Number4000-0105-51
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/16/2021
Initial Date FDA Received05/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-