U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A VENTILATOR
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
 
Event Description
The customer reported to vyaire medical that the 3100a ventilator amplitude is fluctuating all over.At this time, there is no information regarding patient involvement associated with the reported event.
 
Manufacturer Narrative
Results of investigation:the vyaire field service representative (fsr) arrived on site and assessed the device.The fsr bypassed humidifier on device.Successfully preformed circuit calibration and ventilator performance checks, and cleaned rear column filter and replaced o2 inlet filter.The reported issue could not be replicated.It is believed that the power button was in the "locked" position.No issues could be detected in the system at this time.The device was safe and ready to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
MDR Report Key11829142
MDR Text Key250760466
Report Number2021710-2021-13883
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20100414
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A VENTILATOR
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/22/2021
Initial Date FDA Received05/14/2021
Supplement Dates Manufacturer Received05/19/2021
Supplement Dates FDA Received07/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-