Model Number 3100A VENTILATOR |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
|
|
Event Description
|
The customer reported to vyaire medical that the 3100a ventilator amplitude is fluctuating all over.At this time, there is no information regarding patient involvement associated with the reported event.
|
|
Manufacturer Narrative
|
Results of investigation:the vyaire field service representative (fsr) arrived on site and assessed the device.The fsr bypassed humidifier on device.Successfully preformed circuit calibration and ventilator performance checks, and cleaned rear column filter and replaced o2 inlet filter.The reported issue could not be replicated.It is believed that the power button was in the "locked" position.No issues could be detected in the system at this time.The device was safe and ready to use.
|
|