• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACCORD TENSIONER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACCORD TENSIONER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71360020
Device Problem Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2021
Event Type  malfunction  
Event Description
It was reported that during thr surgery, one accord tensioner was slipping the other accord tensioner would not tension above 50.A smith and nephew back up device was available.No injury to patient or significant delay reported.
 
Manufacturer Narrative
H6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device has scratches on the device.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A functional evaluation confirmed that the device was paired with mating part mechanism and does not function as intended.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD TENSIONER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11832475
MDR Text Key250873329
Report Number1020279-2021-04352
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010485359
UDI-Public03596010485359
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360020
Device Catalogue Number71360020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2021
Initial Date Manufacturer Received 04/29/2021
Initial Date FDA Received05/17/2021
Supplement Dates Manufacturer Received06/29/2021
Supplement Dates FDA Received06/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-