Other relevant device(s) are: product id: bi71000483, serial/lot : none.Device evaluated by mfr: a medtronic representative went to the site to perform a system check out and they found that the rad ma was out of tolerance.The left drive wheel was pushed in and rubbing on the side, and the x-stage cover was hitting when extending the gantry causing the system to torque out.As this was found during preventative maintenance, the x-ray, motion, ups, and cmos batteries were replaced, an the system was cleaned and greased.A quote was sent to the site for the x-stage cover.The pendant backlight was also dimming in and out, but still usable.All other functions worked properly.If information is provided in the future, a supplemental report will be issued.
|