Age at time of event: 18 years or older.Device evaluated by manufacturer: returned product consisted of the rotapro atherectomy system.The burr catheter was received attached to the advancer unit.The advancer, handshake connections, sheath, coil, burr and annulus were visually and microscopically examined.Inspection of the device revealed that the sheath was torn at 21.5mm from the distal end of the strain relief.The damage is consistent to the damage caused by over-tightening the hemostasis valve.In order to confirm the reported abnormal noises, the device was functional tested.Functional testing was performed by connecting the rotapro advancer to the rotapro console control system.When the knob (ablation switch) was pressed, the device did not get any speed and stalled.Destructive testing was performed to determine the cause of the device stall, but testing was not able to determine the cause of the stall.Product analysis confirmed the reported damage to the sheath, but the reported unusual noises could not be performed due to the device stalling.Destructive testing was not able to determine the cause of the device stall.
|