Model Number WA47787A
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem Foreign Body In Patient (2687)
Event Date 05/12/2021
Event Type  Injury  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transcervical resection of the fibroids (tcrf) procedure, the loop wire at the distal end of the electrode broke and a fragment fell into the patient.The broken-off fragment could not be retrieved as it was too small to be visible via x-ray.A follow-up procedure was scheduled in one month to complete the procedure.The patient is reportedly fine.
 
Manufacturer Narrative
Additional information: h4 ¿ device manufacturer date.Device evaluation: the suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that the loop wire at the distal end of the hf electrode is broken off due to brittle failure.Thus, the reported event/incident was attributed to component failure.This is a known error pattern, which has already been examined in the past.However, it was assessed that a material change was not required based on the very low number of occurrences of this failure mode.A manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPOINT II BIPOLAR ELECTRODE RESECTOSCOPIC - 4.0 MM ANGLED LOOP, RESECTING
Type of Device
ELECTRODES, ELECTROSURGICAL
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
GYRUS MEDICAL LTD.
fortran road
st. mellons
cardiff
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key11847345
MDR Text Key251353684
Report Number9610773-2021-00140
Device Sequence Number1
Product Code HIN
UDI-Device Identifier14042761086261
UDI-Public14042761086261
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K111751
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA47787A
Device Catalogue NumberWA47787A
Device Lot NumberU2011047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/13/2021
Initial Date FDA Received05/19/2021
Supplement Dates Manufacturer Received01/05/2022
Supplement Dates FDA Received01/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-