• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EPIX THERAPEUTICS DIAMONDTEMP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EPIX THERAPEUTICS DIAMONDTEMP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number CEDT100S
Device Problem No Apparent Adverse Event (3189)
Patient Problem Bronchitis (1752)
Event Date 05/05/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a radiofrequency (rf) ablation procedure, a steam pop occurred.It was also noted that post procedure, the patient had acute bronchitis.Medication was administered and the patient required extended hospitalization.The case was completed with radiofrequency.The patient is a participant in a clinical study.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files and cedt100s radiofrequency catheter with lot 210505080050 were returned and analyzed.The data files showed impedance values were high at the end of ablations.Visual inspection of the distal shaft of the catheter was observed to be bent.The reported steam pop issue could not be replicated under the testing environment.However, the radiofrequency catheter passed electrical test and isolation test.In conclusion, the reported steam pop could not be confirmed through testing.The radiofrequency catheter passed the returned product inspection as per specification.Correction: b3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDTEMP
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
EPIX THERAPEUTICS
945 stewart drive
suite 100
sunnyvale CA 94085
MDR Report Key11848150
MDR Text Key252090676
Report Number3015180993-2021-00010
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
PMA/PMN Number
P200028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/14/2021
Device Model NumberCEDT100S
Device Catalogue NumberCEDT100S
Device Lot Number0000217064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2021
Initial Date Manufacturer Received 05/10/2021
Initial Date FDA Received05/19/2021
Supplement Dates Manufacturer Received06/18/2021
Supplement Dates FDA Received07/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age70 YR
Patient Weight106
-
-