• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL INQUIRY STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL INQUIRY STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-81104
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 05/13/2021
Event Type  Injury  
Event Description
At the end of the procedure, a fibrous thread was observed upon removal of the catheter from the coronary sinus.
 
Manufacturer Narrative
One image was submitted for evaluation to product performance engineering; no device components were returned.The photo appeared to show an inquiry distal tip.A string-like foreign material appeared to be attached to the distal tip.Visual inspection was based solely upon a review of the photograph provided.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY STEERABLE DIAGNOSTIC CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key11855696
MDR Text Key251698537
Report Number3008452825-2021-00303
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05414734302975
UDI-Public05414734302975
Combination Product (y/n)N
PMA/PMN Number
K961924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model NumberIBI-81104
Device Catalogue NumberIBI-81104
Device Lot Number7677434
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/14/2021
Initial Date FDA Received05/20/2021
Supplement Dates Manufacturer Received08/12/2021
Supplement Dates FDA Received08/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
Patient Weight70
-
-