|
Model Number 44021 |
Device Problem
Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A visual inspection of the returned device revealed no visual damages.A functional inspection revealed the telescope assembly was able to properly pull back, advance, and retract.A test guidewire was inserted and came out in the middle of the tip assembly due to a perforation observed in this section.Microscopic inspection revealed the perforation was in the middle section of the tip assembly.
|
|
Event Description
|
Reportable based on device analysis completed on 06may2021.It was reported that crossing issues occurred.The 95% stenosed target lesion was located in a mildly tortuous and moderately calcified superficial femoral artery.An opticross 18 jp imaging catheter was introduced for visualization but was unable to cross the target lesion.The device was removed and the procedure completed with another of the same device.There was no patient injury.However, returned device analysis revealed a device perforation.
|
|
Search Alerts/Recalls
|
|
|