Model Number 1570-01-090 |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
After opening the summit stem, the surgeon noticed some fuzz around the implant and thought it looked odd.Proceeded to open a new stem.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : examination of the returned device is unable to confirm the reported observation.No debris was identified.It is noted that the device not only was transported after having been opened by the complainant it has now been, per process requirement, decontaminated.This may have caused the debris to have detached prior arriving for investigative inspection.The report cannot be determined or a root cause identified.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot l a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|