Model Number 74018911
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka surgery, the bars that acts as spring mechanism on a journey art isrt assm tool broke outside the patient.The procedure was completed, without delay, using the same device.Patient was not harmed.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The returned device has a fractured spring mechanism rendering the device inoperable.The device was manufactured in 2016 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY ART ISRT ASSM TOOL
Type of Device
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11860154
MDR Text Key251840659
Report Number1020279-2021-04539
Device Sequence Number1
Product Code MBH
UDI-Device Identifier03596010557292
UDI-Public03596010557292
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018911
Device Catalogue Number74018911
Device Lot Number16FPG0025
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/06/2021
Initial Date FDA Received05/20/2021
Supplement Dates Manufacturer Received06/11/2021
Supplement Dates FDA Received06/16/2021
Is This a Reprocessed and Reused Single-Use Device? No