Model Number 71175077 |
Device Problems
Connection Problem (2900); Material Deformation (2976); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a trauma surgery, an evos small t8 fixed handle linear driver was worn; hence, it was unable to hold the screw.Surgery was resumed, without any significant delay, with an available smith and nephew back up device; therefore, patient was not harmed.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was not returned for evaluation.Therefore, the product analysis and the reported event could not be confirmed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|