H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport clearvue slim, one catheter, one flushing connector, one introducer needle, one straight non-coring needle, one right-angle non-coring needle, one vein pick, two catheter locks, one syringe, one guidewire in a guidewire hoop, one introducer peel-apart sheath and vessel dilator, one safety infusion set, and one tunneler were returned for evaluation.The investigation is confirmed for the reported deformed package, material deformation, melted, ambient temperature problem as the interior and exterior hard plastic package was found warped during visual evaluation.Furthermore, a crack in the outer tray was also observed.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 10/2021).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|