The customer had two work areas configured in the software and each of them had one validation criteria.In one work area, there was a validation criteria that automatically validates all results received by the cobas infinity software ("validate results always").In the second work area, the validation criteria was to hold and to not automatically validate results that contained any kind of qc flag.The customer had a test that belonged to both work areas, therefore, they could be affected by both validation criteria.During the investigation, it was determined that when a test is contained in two work areas with different validation criteria, one being "validate results always", the cobas infinity will not take into consideration the other criteria, in this case, the qc alarm.The issue has been verified as an issue with the cobas infinity software.
|