Model Number V60
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2021
Event Type  malfunction  
Manufacturer Narrative
Date of report: 27may2021.There was no patient involvement.
 
Event Description
It was reported to philips that the device had an electrical safety test failure.The device was not in clinical use at the time of the event.There was no report of patient or user harm.
 
Manufacturer Narrative
The customer was instructed to securely connect the 2 grounds coming from the front panel assembly towards the grounding plate in front of the power supply to resolve the issue.Upon further investigation, information was received indicating that the reported issue was found during preventative maintenance; there was no patient on the unit.Therefore, this complaint no longer meets reportability requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key11898893
MDR Text Key253128399
Report Number2031642-2021-03875
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Distributor Facility Aware Date04/29/2021
Initial Date Manufacturer Received 04/29/2021
Initial Date FDA Received05/27/2021
Supplement Dates Manufacturer Received06/21/2022
Supplement Dates FDA Received06/22/2022
Date Device Manufactured05/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse