Model Number M004ERFSDS96200 |
Device Problems
Obstruction of Flow (2423); Connection Problem (2900)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
During an ablation procedure an intellanav stablepoint open-irrigated catheter was selected for use.It was reported that the catheter was connected to the maestro generator, however, the generator was unable to recognize the catheter.The generator was reset and the catheter was reconnected.The error persisted.The catheter was then connected to the osypka generator.Flushing difficult was observed as the pump noticed an occlusion in the system.The catheter was exchanged and the procedure was completed with no patient complications.
|
|
Manufacturer Narrative
|
The device was returned to boston scientific for analysis.No visual were noted.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.While manipulating the shaft, continuity checks revealed an electrical open in the ablation thermocouple.The device's lumen was leak tested using an isaac pressure decay test system set to 107 psi, and a touhy bourst (tb) seal for sealing the irrigation holes and mini electrodes (me), the lumen pressure decay was measured three times, with re-seating of the tb seal between each test.Lumen pressure decay values were all below the maximum acceptable limit.The device was connected to a maestro generator and a metriq system then set through critical testing relative to issue.The device was also unexpired and connected to the rhythmia hdx displayed on the workstation.The device appeared on the workstation yet did not appear to be detected on the maestro generator due an open atc.The values of force sensing and direct sense were tested and resulted in normal responsive signals and within the manufacturing standard range.Based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
|
Event Description
|
During an ablation procedure a intellanav stablepoint open-irrigated catheter was selected for use.It was reported that the catheter was connected to the maestro generator, however, the generator was unable to recognize the catheter.The generator was reset and the catheter was reconnected.The error persisted.The catheter was then connected to the osypka generator.Flushing difficult was observed as the pump noticed an occlusion in the system.The catheter was exchanged and the procedure was completed with no patient complications.
|
|
Event Description
|
During an ablation procedure a intellanav stablepoint open-irrigated catheter was selected for use.It was reported that the catheter was connected to the maestro generator, however, the generator was unable to recognize the catheter.The generator was reset and the catheter was reconnected.The error persisted.The catheter was then connected to the osypka generator.Flushing difficult was observed as the pump noticed an occlusion in the system.The catheter was exchanged and the procedure was completed with no patient complications.
|
|
Manufacturer Narrative
|
The device was returned to boston scientific for analysis.No visual were noted.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.While manipulating the shaft, continuity checks revealed an electrical open in the ablation thermocouple.The device's lumen was leak tested using an isaac pressure decay test system set to 107 psi, and a touhy bourst (tb) seal for sealing the irrigation holes and mini electrodes (me), the lumen pressure decay was measured three times, with re-seating of the tb seal between each test.Lumen pressure decay values were all below the maximum acceptable limit.The device was connected to a maestro generator and a metriq system then set through critical testing relative to issue.The device was also unexpired and connected to the rhythmia hdx displayed on the workstation.The device appeared on the workstation yet did not appear to be detected on the maestro generator due an open atc.The values of force sensing and direct sense were tested and resulted in normal responsive signals and within the manufacturing standard range.Based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
|
Search Alerts/Recalls
|