• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problem High impedance (1291)
Patient Problems Nausea (1970); Vomiting (2144)
Event Date 05/27/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for gastric stimulation and gastrointestinal/pelvic floor therapy.It was reported that the patient experienced an increased nausea and vomiting.Upon interrogation, overall impedance is greater than 800 ohms.C<(>&<)>2 >20000, c<(>&<)>3 422 ohms, 2<(>&<)>3 >20000.Parameters changed from 5.7v to 7.0v.Cycling increased to 2s on 3s fa.Previously was 1s on 4s off.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare professional (hcp).The hcp reported that the cause was unknown and that they have modified patient's medication.No further complications were reported/anticipated at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key11913762
MDR Text Key265971450
Report Number3004209178-2021-08598
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2021
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/27/2021
Initial Date FDA Received06/01/2021
Supplement Dates Manufacturer Received06/15/2021
Supplement Dates FDA Received07/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age51 YR
-
-