• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114489
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2021
Event Type  malfunction  
Manufacturer Narrative
Facility name: (b)(6).Phone number: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during set up with a prismaflex machine, a loose wheel was noted.Further inspection observed that the left front wheel and the right rear wheel were damaged.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was evaluated by a qualified technician.It was determined during visual inspection that the left front wheel and the right rear wheel were damaged.To resolve the issues, the qualified technician requested for spare parts for replacement of the damaged wheels.The reported condition was verified.The cause could not be determined.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11914149
MDR Text Key263610618
Report Number9616026-2021-00023
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number114489
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/06/2021
Initial Date FDA Received06/01/2021
Supplement Dates Manufacturer Received06/25/2021
Supplement Dates FDA Received06/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-