|
Model Number 45027 |
Device Problems
Display or Visual Feedback Problem (1184); Difficult to Flush (1251)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device analysis by mfr: the returned product consisted of an angiojet zelante catheter.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The device returned with 300cc of blood in the waste bag.A kink was located 12cm from the tip.Functional testing was attempted.The pump was inserted into the ultra drive unit console.The pump and device would not prime.The devices pressure could not be recorded.The device was dissected at the kinked area and showed a small break in the hypotube.Upon manipulating the shaft to try and get an acceptable photo the hypotube separated due to the damage.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
|
Event Description
|
Reportable based on device analysis result completed on 20may2021.It was reported that the device could not be flushed and an error message appeared.An angiojet zelantedvt catheter was selected for use in a thrombectomy procedure.The device was working fine, until the second time it was used, a check saline supply error message occurred, and the catheter could no longer be flushed.The procedure was completed and there were no patient complications reported.The patient was stable.However, returned device analysis revealed a 12cm kink from the tip and a small break in the hypotube.
|
|
Search Alerts/Recalls
|
|
|