Model Number 550AHCT |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation is in progress, but not yet concluded.
|
|
Event Description
|
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the measured hemoglobin (hgb) value would drop low for no reason.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Updated blocks: h3 and h6 during laboratory analysis, the product surveillance technician (pst) observed the performance of the hematocrit/saturation (h/sat) remained steady throughout evaluation.The complaint indicated volatility in h/sat output.The spring clip was loose and was reattached and then the monitor was put into operate mode.Sample h/sat values were set using an in-vivo adjustment.The system was left in operate mode for four hours and the values were checked again.Values did not significantly change during this time.Sample h/sat on-screen values remained steady for several hours.
|
|
Event Description
|
Per clinical review: according to the information available, the team at the user facility had an incident with the blood parameter monitor (bpm) on (b)(6) 2021 during a cardiopulmonary bypass (cpb) procedure.Additionally, the team had an issue with the bpm not passing its color chip during set up.The reading of the hemoglobin was inaccurate, for it was low compared to the blood gas analyzer.It is not known if the occurrence was before or after a first in-vivo calibration.There was no delay in the continuation of the surgical procedure.There was no harm or blood loss.
|
|
Manufacturer Narrative
|
The reported complaint could not be confirmed.The service repair technician (srt) observed the monitor to pass power on self tests with no errors.The hematocrit saturation (h/sat) probe passed color chip and cuvette testing several times within specification.The reported complaint could not be confirmed.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Search Alerts/Recalls
|