• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Fluid/Blood Leak (1250)
Patient Problem Cough (4457)
Event Date 05/11/2021
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has not been returned to olympus for evaluation.Olympus did help trouble shoot the issue over the phone, and it was discovered that the acecide bottles may have been loaded backwards causing the fluid leak and inability to close the door properly.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Event Description
It is reported that an endoscope reprocessor experienced error code e51 (leak detected).The end user reported that the acecide door initially could not be closed.Both acecide bottles were removed and the drawer could then be closed.At this point it was noted there was a spill of acecide.The end user reported inhaling acecide fumes resulting in coughing and also getting some acecide on this hands.The end user reported wearing proper personal protection equipment.No additional consequences to the end user have been reported.Additional details regarding the end user/reported event have been requested, at this time, no additional information has been provided.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to the user closing the disinfectant bottle drawer with the acecide bottles loaded backward which may have caused the leaking within the equipment.Attempts to requests additional information were completed.However, no customer response was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11921188
MDR Text Key253809244
Report Number8010047-2021-07000
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/11/2021
Initial Date FDA Received06/02/2021
Supplement Dates Manufacturer Received07/27/2021
Supplement Dates FDA Received08/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-